Fluke Calibration 7250 for Rent, Ruska Pressure Controller / Calibrator
Ruska Pressure Controller / Calibrator

Order #: 7250-RENT1

Mfg #: 7250

Questions about availability or need help? Contact Us:

(800) 633-0561
Call for price
Call for price

Fluke Calibration 7250 for Rent, Ruska Pressure Controller / Calibrator

Ruska Pressure Controller / Calibrator

Order #: 7250-RENT1

Mfg #: 7250

Call for price

Request a Rental Quote

Have a Question? Speak to a Transcat Rental Expert: 800.264.4059

Call for price
Availability: Call for Availability
This product is available for shipping to the United States, Canada, and Puerto Rico only.

Product Highlights

  • Pressure ranges from 0 to 5 and 0 to 3 000 psi (0 to 35 kPa and 0 to 21 MPa)
  • Model 7250xi and 7250i provide advanced precision of 0.005 % of reading
  • Model 7250 provides 0.003 % of full scale precision
  • Stability: 0.0075 % of reading per year
  • Time to setpoint: 20 seconds with no overshoot
  • Control stability: 10 ppm
  • Active matrix color screen with enhanced navigation menus
  • Languages: English, French, Chinese, German, Japanese, Spanish and Italian

The Series 7250 Pressure Controllers/Calibrators offers unmatched performance, an active matrix color screen and enhanced control stability. Utilizing multiple sensors and ranges in a single instrument, the Series 7250 combines precision, stability, speed and affordability. Models 7250xi, 7250i and 7250, up to 2,500 psi (17.2 MPa), use a unique quartz sensor, the most accurate pressure sensing technology in a pressure calibrator. Each quartz sensor is manufactured and tested to provide the ultimate performance required by a Fluke Calibration pressure calibrator, ensuring that every customer receives quality, precision and stability in their instrument.

Standard precision
The Fluke Model 7250 pressure calibrator / controller (formerly Ruska 7250) offers an economical approach to automatic pressure testing and calibration with a precision of 0.003 % of full scale for ranges to 2,500 psi (17.2 MPa). A 3,000 psi (21 MPa) full scale range is available with 0.012% of reading precision from 900 psi to 3000 psi (6.3 to 21 MPa) precision.

Models 7250xi and 7250i offer advanced percent of reading precision for increased capability with a single instrument, reducing the investment required to calibrate a wide variety of pressure devices and ranges.

The Model 7250xi provides 0.005 % of reading precision from 5 % to 100 % of the instrument's range. This unmatched precision is achieved through unique quartz pressure sensing technology and multiple quartz sensors in a single instrument. Various ranges from 0 to 20 and 0 to 2 500 psi (0 to 140 kPa and 0 to 17.2 MPa) are available.

The Model 7250i provides 0.005 % of reading from 25 % to 100 % of range and is available in various ranges from 0 to 5 and 0 to 2,500 psi (0 to 35 kPa and 0 to 17.2 MPa).

For pressure below the lower threshold of 5% for the 7250xi and 25 % for the 7250i, the precision becomes 0.005 % of the lower value. For example, a 1,000 psi (7 MPa) Model 7250xi provides 0.005 % of reading from 50 to 1,000 psi (0.035 to 7 MPa); the precision for pressures from 0 to 50 psi (0 to 0.035 MPa) is 0.005 % of 50 psi (0.035 MPa).

High-speed pressure control
All Series 7250 instruments up to 2,500 psi (17.2 MPa) reach setpoint in 20 seconds or less into a 15 cubic inch (245 cc) volume, with no overshoot, to allow high-speed pressure test and calibration.

Dual control modes
The Series 7250 also provides two user-selectable control modes: active and passive mode. In active mode, the 7250 is continually maintaining the setpoint and can compensate for small leaks and pressure changes due to temperature. In passive mode, the user defines a control band and the 7250 will turn off the controller once it achieves the setpoint within the control band. In a leak-free and temperature- stable system, passive mode contributes no additional uncertainty, providing optimum performance.

Automating pressure test and calibration
The 7250 is easy to use and can automate your calibrations in several ways:

Step up/down: for calibrations where the increments are fixed intervals, enter a userdefined step value. The Series 7250 increases or decreases the pressure by the step amount with the jog dial

Win a Generac Generator when purchasing $750 or more from Transcat.Win a Generac Generator when purchasing $750 or more from Transcat.
Pressure Ranges
Model 7250 Any Full Scale pressure range from 5 to 2,500 psig (35 kPa to 17.2 MPa). Or, Permanent Absolute ranges from 15 to 50 psia (100 to 350 kPa). A 3,000 psia range (21 MPa) is also available.
Model 7250i Any Full Scale pressure range from 5 to 2,500 psig (35 kPa to 17.2 MPa). Or, Permanent Absolute ranges from 15 to 50 psia (100 to 350 kPa). A 3,000 psia range (21 MPa) is also available.
Model 7250xi Any Full Scale pressure range from 20 to 2,500 psig (138 kPa to 17.2 MPa).
Optional Modes Absolute using barometric reference sensor for ranges from 15 to 2,500 psig (100 kPa to 17.2 MPa)
Vacuum (negative gauge) for ranges 5 to 2,500 psig (35 kPa to 17.2 MPa)
Vacuum reference (requires external vacuum pump)
Performance
Precision Model 7250
Ranges to 2,500 psi (17.2 MPa): 0.003% of FS
3,000 psi (21 MPa): 0.012% reading from 30% to 100% full scale. Below 30% FS: 0.00036% full scale

Model 7250i
From 25% to 100% full scale (FS): 0.005% reading
Below 25% FS: 0.005% of 25% FS

Model 7250xi
From 5% to 100% full scale (FS): 0.005% reading
Below 5% FS: 0.005% of 5% FS

Precision is defined as the combined effects of linearity, repeatability and hysteresis throughout the operating. Expression of uncertainty conforms with the recommendations of the ISO Guide to the Expression of Uncertainty in Measurement.

Stability Ranges up to 2,500 psi (17.2 MPa): 0.0075% of reading per year
3,000 psia (21 MPa) range: 0.005% of reading
Control Stability Active mode: 0.001% FS (10 ppm) Passive mode: no additional uncertainty
Display Resolution User selectable to 1:1,000,000
Control Response 20 seconds or less with zero overshoot into a 15 in3 volume (245 cc) for ranges to 2,500 psi (17.2 MPa)
60 seconds for 3,000 psia (21 MPa) range
Negative Gauge Precision (optional)

Model 7250
0.003% of FS

Model 7250i
0.005% of 25% FS or 0.00075 psi (0.005 kPa), whichever is greater

Model 7250xi
0.005% of 5% FS or 0.00075 psi (0.005 kPa), whichever is greater

Barometric Reference (optional) 0.002 psi (0.014 kPa) maximum error per year
Vacuum Reference (optional) 0.0002 psi (0.0014 kPa) maximum error per year (Only available on 7250i and 7250)
Calibration
Calibration An accredited calibration report is provided standard.
Total Uncertainty
Maximum deviation from the true value of pressure including precision, stability, temperature effects and the calibration standard
Model 7250 Ranges to 2500 psi (17.2 MPa)
90 day 0.003% FS + 0.002% reading
1 year 0.003% FS + 0.0075% reading

3,000 psi (21 MPa)
30% to 100% FS
90 day: 0.012% reading
1 year: 0.013% reading

Model 7250i Ranges from 25% to 100% full scale
90 day: 0.0055% reading
1 year: 0.009% reading
Model 7250xi Ranges from 5% to 100% full scale
90 day: 0.0055% reading
1 year: 0.009% reading
Control Parameters
Volume 5 to 60 cubic inches (82 to 980 cc)
Low Control (absolute) 0.15 psia (1 kPa)
Communications
Communications RS232 and IEEE-488, SCPI syntax. Series 7215, Model 7010, Series 6000 and DPI 510 emulation are standard.
LabVIEW driver
Firmware updates are performed via RS232 interface.
Languages
Menus and functions The Series 7250 is capable of displaying menus and functions in: English, French, Chinese, German, Japanese, Spanish and Italian.
Options
Options
  • Barometric reference (absolute and vacuum)
  • Vacuum mode (negative gauge)
  • Vacuum reference (requires external vacuum pump)
  • Intecal software
  • NVLAP-accredited calibration
  • Triple scale for 3,000 psia (21 MPa) 7250
  • Rack mount kit
  • Liquid trap assembly
General
Display TFT, VGA, active matrix, 6.4 in (162.5 mm) 640 x 480 resolution, 65,000 colors
Temperature Operating: 64°F to 96°F (18°C to 36°C)
Storage: -4°F to 158°F (-20°C to 70°C)
Humidity 5% to 95% relative humidity, non-condensing
Dimensions (h x w x d) All versions: 7 in x 16.5 in x 19 in (178 mm x 419 mm x 483 mm)
Weight Model 7250/7250i: 17 lb (7.7 kg)
Model 7250xi: 20 lb (9 kg)
Test Port and Supply Connection 1/4 in NPT female
Warm Up Time 2 to 3 hours; may be left on indefinitely
Pressure Medium Nitrogen or clean dry air