Fluke Calibration 5618B-12-P for Rent, Secondary PRT Probe, 12" x 1/8" OD, -200 to 500C
Secondary PRT Probe, 12" x 1/8" OD, -200 to 500 C

Order #: 5618B-12-P-RENT1

Mfg #: 5618B-12-P

Questions about availability or need help? Contact Us:

(800) 633-0561
Call for price
Call for price

Fluke Calibration 5618B-12-P for Rent, Secondary PRT Probe, 12" x 1/8" OD, -200 to 500C

Secondary PRT Probe, 12" x 1/8" OD, -200 to 500 C

Order #: 5618B-12-P-RENT1

Mfg #: 5618B-12-P

Call for price

Request a Rental Quote

Have a Question? Speak to a Transcat Rental Expert: 800.264.4059

Call for price
Availability: Call for Availability
This product is available for shipping to the United States, Canada, and Puerto Rico only.

Product Highlights

305 mm (12 in) Small Diameter Probe, lead wire (cable) 6 ft

(17025 accredited calibration included. Traceable to NIST standards.)

X = termination. Specify "A" (INFO-CON for 914X), "B" (bare wire), "D" (5-pin DIN for Tweener Thermometers), "G" (gold pins), "J" (banana plugs), "L" (mini spade lugs), "M" (mini banana plugs), "P" (INFO-CON for 1523 or 1524), or "S" (spade lugs).

Secondary level performance with full ITS-90 calibration

  • Small diameter sheath, 3.2mm (0.125 in)
  • Excellent stability
  • Includes ITS-90 coefficients
  • NVLAP-accredited calibrated from -200°C to 500°C, lab code 200706-0

For secondary level performance with full ITS-90 calibration, Fluke Calibration's 5618B series PRTs are an excellent choice for critical temperature measurements. Featuring a 3.2 mm diameter (1/8 in) sheath, these industrial standards probes have reduced response time without compromising precision. This small diameter 5618A probe works well in many applications where immersion depth is limited. Larger diameter probes give more measurement error in short immersion depth applications because they conduct more heat between ambient and the sensor.

With each probe you will receive a full NVLAP-accredited calibration report. On the report you'll get the test data and the ITS-90 calibration coefficients that you can easily input into your Fluke Calibration thermometer.

The 5618B is also a great probe to use for calibrating your Fluke Calibration 9132 or 9133 infrared calibrators. In fact, these IR black body heat sources were designed to be calibrated with this type of probe. Now you can calibrate these targets in your own lab!

For use from -200 °C to 500 °C (the six-inch model goes to 300 °C), you won't find a better industrial standard in this configuration than our 5618B. We recommend using the 5618B PRTs with the 1523, 1524, 1502A, 1529, or 1560 thermometer readouts.

Win a Generac Generator when purchasing $750 or more from Transcat.Win a Generac Generator when purchasing $750 or more from Transcat.
Resistance Nominal 100 Ω at 0 °C
Temperature Coefficient 0.003923 Ω/Ω/ °C nominal
Temperature Range -200 °C to 500 °C (-200 °C to 300 °C for 5618B-6-X)
Drift Rate ± 0.1 °C when used periodically to 500 °C
Sheath Material 316 SST
Leads 22 AWG PTFE, 6'
Termination Specify
Hysteresis Less than 0.01 °C at 0 °C when using -196 °C and 420 °C as the end points.
Time Constant Four seconds maximum for 63.2 % response to step change in water moving at 3 fps.
Thermal EMF Less than 25 mV at 420 °C
Calibration Includes manufacturer's NVLAP-accredited calibration w/ITS-90 coefficients, R vs. T values in 1 °C increments, lab code 200706-0
Size
  • 5618B-12: 305 mm L x 3.2 mm dia. (12 x 1/8 in)
  • 5618B-9: 229 mm L x 3.2 mm dia. (9 x 1/8 in)
  • 5618B-6: 152 mm L x 3.2 mm dia. (6 x 1/8 in)
Calibrated Accuracy (k=2)
  • ± 0.026 °C at -200 °C
  • ± 0.066 °C at 0 °C
  • ± 0.219 °C at 500 °C