Overview

Product Highlights

  • Maximum pressure control range up to 15,000 psi (1,000 Bar)
  • Interchangeable pressure module
  • Available in oil and water versions
  • Accuracy of 0.02%FS, 0.01%FS, or 0.01%RD
  • Dual-range from -15~3,000 psi (-1~200 Bar) to 0~15,000 psi (0~1,000 Bar)
  • Control stability of 0.005%FS
  • Ultra-High speed pressure generation and control
  • No external pressure source required
  • Control using internal or external pressure modules
  • Contamination Prevention System (CPS) (ADT793 optional)
  • LAN, USB, RS232, and Ethernet communication
  • External reservoir support
  • Easy maintenance
  • Emulation mode
  • Pressure modules are not included

Technical Resources

Specifications

Specification ADT793
Pressure Range 0 to 15,000 psi (1,000 Bar)
Precision[1] 0.008%RD or 0.007%FS or 0.01%FS
Accuracy[2] 0.01%RD or 0.01%FS or 0.02%FS
Control Stability[3] < 0.005%FS, typically 0.003%FS
Control Response Time[4] < 30 Seconds
Pressure Type[5] Gauge, absolute
Internal Pressure Control Modules 1
External Pressure Control Modules 1
Max Pressure Range of Internal Module 0~15,000 psi (0~1,000 Bar)
Min Pressure Range of Internal Module -15~3,000 psi (-1~200 Bar)
Min Range of External Module -15~1,000 psi (-1~70 Bar)
Range Switching Mode[6] Fixed, auto
Pump Source Type[7] Built-in hydraulic pump, no external pressure source required.
Control Mode Fast, standard, custom
Maximum Overshoot < 1%FS
Maximum Load Volume < 10,000 psi: Max: 80 ml, 50 mL recommend
10,000~15,000 psi: Max: 60 ml, 35 mL recommend
Contamination Prevention System (CPS) Optional
Pressure Port 1/4 BSP M
Built-in Reservoir Volume 800 mL
External Reservoir Extension Optional
Port Filter[8] Support
Media Diethylhexyl Sebacate or deionized water