MPAC-V Acoustic Camera Verifier/Ultrasound Generator for use with MPAC208 Acoustic Camera, acting as a portable ultrasound generator and verifier to confirm the camera's functionality by creating known ultrasonic signals, allowing technicians to verify leak/PD detection accuracy in the field, ensuring reliable inspection of gas leaks, partial discharge (corona, arcing) in electrical systems, and mechanical issues from a distance. It helps validate the camera's microphone array and processing, ensuring it's calibrated for precise, safe fault finding in industrial environments, even in hazardous ATEX zones.
- Enables the user to perform a dedicated verification procedure on the MPAC camera, which checks key operating parameters in the field.
- Ensures the optimal and reliable performance of the camera's 128 or 208 channel MEMS microphone array by confirming its frequency response, SPL, and microphone alignment are within specification.
- The camera's self-check function, in conjunction with the MPAC-V, helps verify the integrity of the internal components and guarantees consistent performance.
- The MPAC-V verifier/ultrasound generator is not certified for use in explosive atmospheres, unlike some versions of the main camera (e.g., MPAC128-ATEX, MPAC208-IECEx).
Applications
- Megger MPAC128 Acoustic Imager
Each Unit Includes
- MPAC-V Acoustic Camera Verifier/Ultrasound Generator